© National Instruments Corporation 7 NI PXI/PCI-5114 Specifications
Phase-Locked Loop (PLL) Reference Clock
External Sample Clock
Sources CLK IN (front panel SMB connector)
Frequency
Range
50 MHz to 250 MHz Divide by n
decimation available
where
1 n 65,535.
For more information
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
Duty Cycle
Tolerance
45% to 55%
Specification Value Comments
Sources NI PXI-5114 NI PCI-5114
PXI_CLK10 (backplane
connector
)
CLK IN (front panel SMB
connector)
RTSI 7
CLK IN (front panel SMB
connector)
Frequency
Range
1 MHz to 20 MHz in 1 MHz increments.
Default of 10 MHz.
The PLL Reference Clock frequency must be accurate
to ±50 ppm.
Duty Cycle
Tolerance
45% to 55%
Exported
Reference Clock
Destinations
NI PXI-5114 NI PCI-5114
PFI <0..1> (front panel
9-pin mini-circular
DIN connector)
PXI_Trig <0..7>
(backplane connector)
PFI <0..1> (front panel
9-pin mini-circular
DIN connector)
RTSI <0..7>
Specification Val ue Comments