National Instruments NI PXI/PCI-5114 Bedienungsanleitung Seite 6

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 20
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 5
NI PXI/PCI-5114 Specifications 6 ni.com
Horizontal
Sample Clock
Specification Val ue Comments
Sources Internal, Onboard Clock (internal VCXO)*
External, CLK IN (front panel SMB connector)
*
Internal Sample
Clock is locked to the
Reference Clock or
derived from the
onboard VCXO.
Onboard Clock (Internal VCXO)
Sample Rate
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
*
Divide by n
decimation used
for all rates less than
250 MS/s. For more
information about
Sample Clock and
decimation, refer to
the NI High-Speed
Digitizers Help.
3.815 kS/s to 250MS/s* 250 MS/s to 5 GS/s in
increments of 250 MS/s
Timebase
Frequency
250 MHz When not using
External Sample
Clock.
Timebase
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
±25 ppm Equal to the Reference
Clock accuracy
Sample Clock
Delay Range
±1 Sample Clock period
Sample Clock
Delay Resolution
20 ps
Seitenansicht 5
1 2 3 4 5 6 7 8 9 10 11 ... 19 20

Kommentare zu diesen Handbüchern

Keine Kommentare