National-instruments Isolated Analog Output Module SCC-AO10 Bedienungsanleitung Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 12
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
© National Instruments Corporation 7 SCC-AO10 Isolated Analog Output Module User Guide
5. Adjust the offset potentiometer on the top of the SCC-AO10 until the
voltmeter reads 0 VDC ±100 μV. The offset is now calibrated.
6. Using your software, have AO (X) output 9.9 VDC.
Adjust the gain potentiometer on the top of the SCC-AO10 until the
voltmeter reads 9.9 VDC. The gain is now calibrated.
Specifications
These ratings are typical at 25 °C unless otherwise stated.
Electrical
Number of output channels .................... One nonreferenced (floating)
Input range ............................................. ±10 V
Output range........................................... ±10 V
Current drive .......................................... ±30 mA
Gain nonlinearity.................................... 0.5% of full-scale output range
Output noise ........................................... 2.5 mV
rms
typ; 4 mV
rms
max
Bandwidth .............................................. >23 kHz
Slew rate................................................. 1 V/μs
Short circuit protection
Current ............................................ 37 mA max
Duration .......................................... Unlimited
Load impedance ..................................... 10,000 pF max