National-instruments Isolated Analog Input Modules SCC-AI03 Bedienungsanleitung Seite 6

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 14
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 5
SCC-AI Series Isolated Analog Input Modules User Guide 6 ni.com
Figure 2. Nonreferenced Signal Connection for the SCC-AI (One Channel)
If a high common-mode voltage is present, connect the negative input pins,
1 and 3, to the signal reference. Figure 3 shows a ground-referenced signal
connection on one channel of the SCC-AI.
Figure 3. Ground-Referenced Signal Connection for the SCC-AI
with High Common-Mode Voltage (One Channel)
For information about how to configure the SCC-AI module using
NI-DAQmx, refer to the SCC Quick Start Guide.
+
4
Lowpass
Filter
+
+
Lowpass
Filter
3
2
1
E/M Series DAQ Device
Signal
Source
AI (X )
SCC-AI
AI SENSE
AI GND
AI (X+8)
AI SENSE on
SC-2345/2350
terminal block
or
AI GND on
SCC-68
10 to 100 kΩ
+
4
Lowpass
Filter
+
+
Lowpass
Filter
3
2
1
E/M Series DAQ Device
Signal
Source
AI (X )
SCC-AI
AI SENSE
AI GND
AI (X+8)
High
CMV
+
Seitenansicht 5
1 2 3 4 5 6 7 8 9 10 11 12 13 14

Kommentare zu diesen Handbüchern

Keine Kommentare