CALIBRATION PROCEDURE
SCXI -1125
Introduction
This document contains information and step-by-step instructions for
verifying and calibrating the National Instruments SCXI-1125 signal
conditioning module.
What Is Calibration?
Calibration is a procedure of reading offset and gain errors from a device
and updating special analog calibration circuitry that corrects these errors.
National Instruments calibrates every SCXI-1125 module at the factory.
During the factory calibration process, the calibration constants are stored
in the non-volatile memory of the device. These values are loaded from
memory and used as needed by the device.
Why Should You Calibrate?
Offset and gain errors drift with time and temperature changes, which could
invalidate the factory-set calibration of a device and increase measurement
uncertainty. Calibration restores the device to its specified accuracy.
How Often Should You Calibrate?
The measurement accuracy requirements of your application
determine how often you should calibrate your SCXI-1125 module.
National Instruments recommends you perform a complete calibration
at least once every year. You can shorten this interval to six months or
90 days, based on the demands of your application.
Seitenansicht 0
1 2 3 4 5 6 ... 13 14

Inhaltsverzeichnis

Seite 1 - SCXI -1125

CVI™, LabVIEW™, National Instruments™,ni.com™,NI-DAQ™, and SCXI™are trademarks of National Instruments Corporation. Product and companynames mentioned

Seite 2 - Software

SCXI-1125 Calibration Procedure 10 ni.comVerifying Adjusted ValuesAfter completing the adjustments to your SCXI-1125 module, it isimportant that you v

Seite 3 - Calibration

© National Instruments Corporation 11 SCXI-1125 Calibration ProcedureFront and Rear Signal Connector DiagramsFigure 1. Front Signal Connector Pin Assi

Seite 4 - Setting Up Your Module

SCXI-1125 Calibration Procedure 12 ni.comFigure 2. Rear Signal Connector Pin AssignmentsSERCLKDIGGNDSLOT0SEL*DAQD*/ASERDATINOUTREFMCH7+MCH6+MCH5+MCH4+

Seite 5 - SCXI_Single_Chan_Setup

© National Instruments Corporation 13 SCXI-1125 Calibration ProcedureSpecificationsThe following table contains the specifications you should use when

Seite 6 - Calibrate_E_Series

SCXI-1125 Calibration Procedure 14 ni.com200 –0.025 –0.024750 –0.02468642 –0.02481358250 0.02 0.019800 0.01985126 0.01974874250 0 0.000000 0.000022 –0

Seite 7

SCXI-1125 Calibration Procedure 2 ni.comEquipment and Other Test RequirementsThis section describes the equipment, software, documentation, and testco

Seite 8

© National Instruments Corporation 3 SCXI-1125 Calibration ProcedureDocumentationThe NI-DAQ Function Reference Online Help file and NI-DAQ UserManual

Seite 9 - SCXI_Cal_Constants

SCXI-1125 Calibration Procedure 4 ni.comSetting Up Your ModuleComplete the following steps to set up your module for testing:1. Install the SCXI-1125

Seite 10 - Verifying Adjusted Values

© National Instruments Corporation 5 SCXI-1125 Calibration Procedure7. Connect the calibrator to the appropriate analog input, startingwith channel 0.

Seite 11

SCXI-1125 Calibration Procedure 6 ni.com13. CallSCXI_Scale. Set the following parameters:• SCXIChassisID—The value assigned by MAX• moduleSlot—1• chan

Seite 12

© National Instruments Corporation 7 SCXI-1125 Calibration Procedure• calOp—ND_SELF_CALIBRATE• setOfCalConstants—ND_USER_EEPROM_AREA• calRefVolts—0.04

Seite 13 - Specifications

SCXI-1125 Calibration Procedure 8 ni.com• filterMode—1• cutoffFrequency—10,0009. Connect the calibrator to the analog input you want to adjust. Refert

Seite 14

© National Instruments Corporation 9 SCXI-1125 Calibration Procedure16. Record binary1 with volt1 and binary2 with volt2 for later use.17. Repeat step